        Priming Filter

        Many medical devices require vents to keep their internal pressure in balance with the ambient pressure. This avoids the deformation of rigid plastic containers and interruptions in flow of fluids.

        Our product must protect the fluid path from bacterial contamination (must maintain sterility).

        One good example of this is maintaining sterility of an IV set and glass solution bottle, and to equalize the internal pressure of the bottle via a vent on the spike. Venting characteristics can be obtained by using hydrophobic membranes.

        GVS has a wide range of air vent products that incorporate PTFE or hydrophobic acrylic membrane for this function.

        Download document

        Ask for information

        Properties

        GVS air vents are manufactured with fully automatic over-molding technology that guarantees superior quality in manufacturing parts with high cavity molds.

        Because we use many different raw materials for the frames and/or housings of our products, it is possible to choose assembly methods of the mechanical sealing (interference fit) or bonding (standard with cyclohexanone, methyl cyanoacrylate, or methyl ethyl ketone).

        Available air vents have female Luer locks or male Luer slip connectors.

        Characteristics

        Ref. Standard
        ISO 8536-4
        Efficiency of Filtration
        100%
        Pyrogenicity
        < 0.25 EU/ml
        Maximum Operating Temperature
        50°C

        Applications

        • Infusion

        Product Variants

        Multiple versions of Priming Filter are available.

        Product Variants

        4 available results

        1. TA225BV29G002D01
          TA225 - Priming Filter In PP/0.2 Μm Acrylic - Blue
          Air flow
          >10 ml/min @153cmH2O
          Assembly method
          interference
          Bacterial retention
          "brevundimonas diminuta; "
          Box quantity
          50000
          Effective filtration area (cm²)
          0.107
          Filter material
          Membrane Hydrophobic Acrylic Copolymer
          Filtration Efficiency
          100%
          Housing material
          PP
          Outside diameter (mm)
          Female Luer Lock
          Pore size (µm)
          0.2
          Pyrogenicity EU/ml
          <0.25
          Ref. standard
          ISO 8536-4 / ISO 594-1/2
          Temperature
          50°C
          Total height (mm)
          18.6
          Water breakthrough
          0.5 bar for 5 minutes
        2. TA225BV29G008E01
          TA225 - Priming filter in PP/0.8 µm PVDF - blue
          Filtration Efficiency
          100%
          Air flow
          11 ml/min @20 cmH2O
          Assembly method
          interference
          Box quantity
          50000
          Effective filtration area (cm²)
          0.107
          Filter material
          Membrane Hydrophobic Acrylic Copolymer, PVDF
          Housing material
          PP
          Outside diameter (mm)
          Female Luer Lock
          Pore size (µm)
          0.8
          Pyrogenicity EU/ml
          <0.25
          Ref. standard
          ISO 8536-4
          Temperature
          50°C
          Total height (mm)
          18.6
          Water breakthrough
          0.3 bar for 1 minute
        3. TA225BV29G012D01
          TA225 - Priming Filter In PP/1.2 Μm Acrylic - Blue
          Filtration Efficiency
          100%
          Air flow
          40 ml/min @ 20cm H20
          Assembly method
          interference
          Bacterial retention
          candida albicans
          Box quantity
          50000
          Effective filtration area (cm²)
          0.107
          Filter material
          Membrane Hydrophobic Acrylic Copolymer
          Housing material
          PP
          Outside diameter (mm)
          Female Luer Lock
          Pore size (µm)
          1.2
          Pyrogenicity EU/ml
          <0.25
          Ref. standard
          ISO 8536-4
          Temperature
          50°C
          Total height (mm)
          18.6
          Water breakthrough
          0.177 bar for 15 minutes
        4. TA225BV29G012E01
          TA225 - Priming Filter In PP/3.0 Μm Acrylic - Blue
          Filtration Efficiency
          100%
          Assembly method
          interference
          Bacterial retention
          candida albicans
          Box quantity
          50000
          Effective filtration area (cm²)
          0.107
          Filter material
          Membrane Hydrophobic Acrylic Copolymer
          Housing material
          PP
          Notes
          * Suitable for any type of Male Luer Lock, New Rotating Male Luer Lock and Male Luer Slip
          Outside diameter (mm)
          Female Luer Lock
          Pore size (µm)
          3
          Pyrogenicity EU/ml
          <0.25
          Ref. standard
          ISO 8536-4
          Temperature
          50°C
          Total height (mm)
          18.6