To access the localized content, select the website associated with the Division of your interest:

Health & Safety

    Healthcare & Life Sciences

      Energy & Mobility

        • Priming Purge Filter, image 1

        Priming Purge Filter

        TA225

        Many medical devices require vents to keep their internal pressure balanced with the ambient pressure.

        This avoids deformation of rigid plastic containers and interruptions in flow of fluids.

        Our product must protect the fluid path from bacterial contamination (must maintain sterility).

        One good example of this is maintaining sterility of an IV set and glass solution bottle, and to equalize the internal pressure of the bottle via a vent on the spike.

        TDS

        Ask for information

        Properties

        Venting characteristics can be obtained by using hydrophobic membranes.

        GVS has a wide range of air vent products that incorporate PTFE or hydrophobic acrylic membrane for this function.

        Characteristics and Advantages

        Effective filtration area (cm2)
        0,107
        Temperature
        50°
        Filtration efficiency
        100%
        Pyrogenicity
        <0,25 EU/ml

        Applications

        • Infusion

        Product Variants

        Multiple versions of Priming Purge Filter are available.

        Product Variants

        4 available results

        1. TA225BV29G002D01
          Priming Purge filter TA225 in PP/0.2 µm acrylic - blue
          Air flow
          >10 ml/min @153cmH2O
          Assembly method
          interference
          Bacterial retention
          "brevundimonas diminuta; "
          Box dimension
          60x40x25 cm
          Box quantity
          50000
          Box weight
          17.4 kg
          Color
          Blue
          Effective filtration area (cm²)
          0.107
          Filter material
          Membrane Hydrophobic Acrylic Copolymer
          Filtration efficiency
          100%
          Housing material
          PP
          Notes
          * Suitable for any type of Male Luer Lock, New Rotating Male Luer Lock and Male Luer Slip
          Outside diameter (mm)
          Female Luer Lock
          Pore size (µm)
          0.2
          Pyrogenicity
          <0.25 EU/ml
          Ref. standard
          ISO 8536-4 / ISO 594-1/2
          Temperature
          50°C
          Total height (mm)
          18.6
          Water breakthrough
          0.5 bar for 5 minutes
        2. TA225BV29G008D01
          Priming Purge filter TA225 in PP/0.8 µm PVDF - blue
          Filtration efficiency (%)
          100%
          Air flow
          11 ml/min @20 cmH2O
          Assembly method
          interference
          Box dimension
          60x40x25 cm
          Box quantity
          50000
          Box weight
          17.4 kg
          Color
          Blue
          Effective filtration area (cm²)
          0.107
          Filter material
          Membrane Hydrophobic Acrylic Copolymer, PVDF
          Filtration efficiency
          100%
          Housing material
          PP
          Notes
          * Suitable for any type of Male Luer Lock, New Rotating Male Luer Lock and Male Luer Slip
          Outside diameter (mm)
          Female Luer Lock
          Pore size (µm)
          0.8
          Pyrogenicity
          <0.25 EU/ml
          Ref. standard
          ISO 8536-4
          Temperature
          50°C
          Total height (mm)
          18.6
          Water breakthrough
          0.3 bar for 1 minute
        3. TA225BV29G012D01
          Priming Purge filter TA225 in PP/1.2 µm acrylic -blue
          Filtration efficiency (%)
          100%
          Air flow
          40 ml/min @ 20cm H20
          Assembly method
          interference
          Bacterial retention
          candida albicans
          Box dimension
          60x40x25 cm
          Box quantity
          50000
          Box weight
          17.4 kg
          Color
          Blue
          Effective filtration area (cm²)
          0.107
          Filter material
          Membrane Hydrophobic Acrylic Copolymer
          Filtration efficiency
          100%
          Housing material
          PP
          Notes
          * Suitable for any type of Male Luer Lock, New Rotating Male Luer Lock and Male Luer Slip
          Outside diameter (mm)
          Female Luer Lock
          Pore size (µm)
          1.2
          Pyrogenicity
          <0.25 EU/ml
          Ref. standard
          ISO 8536-4
          Temperature
          50°C
          Total height (mm)
          18.6
          Water breakthrough
          0.177 bar for 15 minutes
        4. TA225R030D01
          Priming Purge filter TA225 in PP/3.0 µm acrylic -blue
          Filtration efficiency (%)
          100%
          Assembly method
          interference
          Bacterial retention
          candida albicans
          Box dimension
          60x40x25 cm
          Box quantity
          50000
          Box weight
          17.4 kg
          Color
          Blue
          Effective filtration area (cm²)
          0.107
          Filter material
          Membrane Hydrophobic Acrylic Copolymer
          Filtration efficiency
          100%
          Housing material
          PP
          Notes
          * Suitable for any type of Male Luer Lock, New Rotating Male Luer Lock and Male Luer Slip
          Outside diameter (mm)
          Female Luer Lock
          Pore size (µm)
          3
          Pyrogenicity
          <0.25 EU/ml
          Ref. standard
          ISO 8536-4
          Temperature
          50°C
          Total height (mm)
          18.6

        This website uses first-party and third-party cookies in order to offer dynamic content and a better user experience. For further information please read our Privacy Policy.