        TP010

        Transducer protectors (TP) are used in hemodialysis blood lines to keep the blood side of the circuit separated from the machine side, and to prevent contamination of the machine by the blood flowing through the circuit.
        This contamination could be very dangerous and can lead to patient cross contamination with hepatitis B or other virus. Transducer protectors let air pass while preventing the blood from passing through. TP’s incorporate bacterial retentive membrane (0.2μm, if not otherwise requested).

        Transducer protectors must guarantee a sufficient air flow to provide immediate reaction time if there is a change in patient blood pressure. If the TP does not react quickly enough, this can be very dangerous for the patient.

        TDS

        Ask for information

        Property

        GVS transducer protectors are also suitable as air or gas vents. Venting characteristics are achieved by using hydrophobic membranes. GVS has a full range of transducer protector products which incorporate either highly hydrophobic PTFE or hydrophobic acrylic membranes.

        GVS transducer protectors are manufactured with fully automated assembly technology, and the sealing is done without any glues or solvents – they are ultrasonically welded. Automatic testing machines keep the manufacturing process consistent and guarantee superior quality.

        The wide range of available media (PTFE or acrylic membranes) provides high flexibility and covers almost any requirement of the market.

        Because of our wide choice of shapes (Flat TP, TP "bell" shape and TP "dome") and the very different type of materials and media used, the GVS transducer protector line is suitable in any blood line or other medical device.

        Available membrane sizes are from 0.02 to 10 micron. Pressure isolator with flexibe diaphragm are available for open heart surgery.

        Characteristics

        Dimensions
        width: 38 mm; length: 60.72 mm
        Housing material
        Rigid PVC / PVC DEHP Free
        Filter material
        flexible PVC film
        Inlet / Outlet Connections
        MLS Male Luer Slip
        Ref. Standard
        ISO 1135-4
        Assembly Method
        bonding

        Applications

        • Open heart surgery

        Product Variants

        Multiple versions of TP010 are available.

        Product Variants

        2 available results

        1. TP010AAT0000AA00
          TP10 Isolator / 1IN+1OUT
          Assembly method
          bonding
          Box dimension
          60x40x35 cm
          Box quantity
          1.000
          Box weight
          11.8 kg
          Color
          clear
          Filter material
          Flexible PVC film
          Housing material
          Rigid PVC
          Inlet / outlet
          MLS Male Luer Slip
          Length (mm)
          60.72
          Maximum operating temperature
          50°C
          Pressure (bar)
          6 bar
          Pyrogenicity
          <0.25 EU/ml
          Ref. standard
          ISO 1135-4 / 10993-1
          Weight (g)
          9.7
          Width (mm)
          38
        2. TP010ARA0000AA00
          TP10 Isolator PVC DEHP Free / 1 Inlet and 1 Outlet
          Assembly method
          bonding
          Box dimension
          60x40x35 cm
          Box quantity
          1.000
          Box weight
          11.8 kg
          Color
          clear
          Filter material
          Flexible PVC film
          Housing material
          PVC DEHP Free
          Inlet / outlet
          MLS Male Luer Slip
          Length (mm)
          60.72
          Maximum operating temperature
          50°C
          Pressure (bar)
          6 bar
          Pyrogenicity
          <0.25 EU/ml
          Ref. standard
          ISO 1135-4 / 10993-1
          Weight (g)
          9.7
          Width (mm)
          38