To access the localized content, select the website associated with the Division of your interest:

Health & Safety

    Healthcare & Life Sciences

      Energy & Mobility

        • TP031, image 1

        TP031

        GVS transducer protectors are manufactured using fully automated assembly technology, and sealing is done without glue or solvents - they are ultrasonically welded. Automatic control machines keep the production process constant and guarantee superior quality.

        Ask for information

        Property

        • Wide adaptability on haemodialysis machines
        • Stability after gamma or beta irradiation
        • High assurance of no leakage
        • No deformation of parts after storage
        • Bacterial and viral barrier

        Characteristics

        Dimensions
        Width: 27.40 mm Length: 26.60 mm
        Effective Filtration Area
        3.6 cm2
        Weight
        2.5 g
        Pressure
        3.2 bar
        Maximum Operating Temperature
        50°C
        Pyrogenicity
        <0.25 EU/ml

        Applications

        • Breathing systems
        • Hemodialysis

        Product Variants

        Multiple versions of TP031 are available.

        Product Variants

        1 available result

        1. TP031ANAG002AA01
          Transducer Protector TP31 - PVC/0.2 µm acrylic - clear/clear
          Air flow
          >15 ml/min
          Assembly method
          bonding
          Bacterial retention
          Brevundimonas diminuta
          Box dimension
          60x40x25 cm
          Box quantity
          5.000
          Box weight
          13.5 kg
          Color
          clear/clear
          Effective filtration area (cm²)
          3.6
          Filter material
          Acrylic Copolymer
          Filtration efficiency
          100%
          Housing material
          PVC
          Inlet / outlet
          "Inlet: MLS Male Luer Slip Outlet: MLS Male Luer Slip + Tube"
          Length (mm)
          26.6
          Maximum operating temperature
          50°C
          Pore size (µm)
          0.2
          Pressure (bar)
          3.2 bar
          Pyrogenicity
          <0.25 EU/ml
          Ref. standard
          ISO 1135-4 / 8638 / 10993-1
          Water breakthrough
          1 bar/1 minute
          Weight (g)
          2.5
          Width (mm)
          27.4